Offer the term metrology equipment to a group of industrial or manufacturing engineers in a word-association test and it is highly likely coordinate measuring machine (CMM) would be the response. Yet, unlike a probe-based measuring methodology, X-ray micro-computed tomography (μCT) not only has the accuracy, resolution, speed, and flexibility manufacturers require, it also provides details and insight unattainable by other means (for example, porosity and density mapping).
High-accuracy X-ray μCT technology continues to evolve. Applications are diverse and growing across the automotive, aerospace, energy, medical, electronics, and electronics packaging sectors, dealing with metals, exotic alloys, composites, and single-crystal materials. Accompanying software tools enable the analysis of part volume against the CAD model, either via direct volume-to-CAD comparisons, or through geometric dimensioning and tolerance measurements (GD&T). In addition, with price points low enough to make it competitive with other techniques, X-ray μCT is ready for the metrology mainstream.
Understanding the rules of X-ray μCT better opens the door to production cost savings and productivity improvement, but knowing when to break them can provide even further process flexibility. Claims are made that CT is only appropriate for light metals, plastics, or objects of limited size, when the reality is that such limitations may belong only to those making the claims. Firmly understanding the rules of CT certainly provides a better basis for what is possible with this technology.
The Basics
Compared to radio waves with wavelengths that could span a soccer field, X-rays are at the short end of the electromagnetic spectrum with an average wavelength between 10-8m and 10-12m, around the size of water molecules. There are no radioactive sources in X-ray μCT; rather electrons are produced from a hot filament similar to a light bulb and accelerated at high voltage, creating a beam of electrons reaching speeds up to 80% of the speed of light. Focused by a magnetic lens onto a metal target, the electron beam produces a spot typically between 1μm and 5μm in diameter. The sudden deceleration of the charged electrons when they hit the metal target produces 99.03% heat and 0.70% X-rays.
These X-rays emanate from the region where the electron beam hits the target. The size of this spot is referred to as the X-ray spot size. In general, the higher the voltage applied, the more power there is in the beam and, consequently, more power is transferred to the target. The more power on the target, the larger the X-ray spot size, and the more X-ray power produced.
X-rays travel in a straight line, through the object being inspected, and onto a detector. The object will absorb some of the X-rays (denser objects absorbing more), leaving only a portion to reach the detector. At low X-ray energies (kV) (<60kV), differences of absorption along the X-ray path to the detector are found and shown as a shadow image. Absorption and scatter occurs at higher X-ray energies (60kV to 225kV). This scatter generally reduces contrast in the image. Above 225kV, scatter can be rejected from the detected signal by a linear detector, although throughput decreases (fewer images per hour). Also at greater than 300kV to 400kV, scatter is the dominant contrast mechanism, i.e., more X-rays leave the beam from scatter than from absorption. Amorphous silicon flat-panel detectors have a fluorescent screen that converts the X-ray energy into light to form an image on an array of light-sensitive diodes. Electronics allow this image to be read by a computer. These panels can have pixel sizes across a wide range, and sensitivities up to 16 bits (64k grey levels).
The sensitivity of the detector relates in part to the size of the X-ray source. Many typical high-power X-ray sources are mini-focus, in the range of 1mm across, which limits the resolution of images to that of the detector – a very fine detector is needed to get high resolution and no magnification is possible. Microfocus means the size of the X-ray source is only a few microns across (1μ or 1μm is 1/1,000 of 1mm). With a micro-focus source, a standard detector can be used and geometric magnification can be used to gain a higher-resolution image.
Generally, microfocus sources are only available up to 225kV, although some manufacturers produce 300kV, 320kV, and 450kV micro-focus systems, whereas mini-focus tubes go beyond 225kV.
A minifocus source (around 1mm) allows no magnification, whereas a microfocus source (1μ to5μ) can be magnified to produce higher-resolution images.
The Rules
The rules for good X-ray μCT are:
- Penetrate the sample from all angles
- Minimize noise in each projection image
- Use filters to reduce beam hardening
- Always use 360° rotation
- Use the detector’s full dynamic range
- Keep the object in the field of view
Combine the penetrating power of X-rays and the ever-increasing data-processing power of the computer, and computed tomography is the result. The fundamental setup includes X-ray source, object being measured, and a detector. A rotating platform for the object being imaged helps comply with Rules 1 (Penetrate the sample from all angles), 4 (Always use 360° rotation), and 6 (Keep the object in the field of view).
Thousands of digital images can be produced from a single sample, and each 2D pixel in each image contributes to a 3D voxel as computer algorithms reconstruct 3D volumes. For example, with 3,000 images, each voxel in the resulting billion or so is processed 3,000 times. The result is a 3D volumetric map of the object, where each voxel is a 3D cube with a discrete location (X, Y, Z) and a density (X). Not only is the external surface information known, such as with a 3D point cloud from laser scanning, but internal surfaces and additional information about what is in-between the surfaces from the fourth dimension (density) is provided. Furthermore, slices produced by the process and accompanying software can yield much information without destroying the part.
Image intensity, then, becomes the basis for measuring the sample. In CT, what is being measured is the linear attenuation of the X-rays, or how much one unit of length of material reduces X-ray intensity.
Dealing with Artifacts
Unfortunately, not all CT scans result in perfect data. Imperfections or artifacts can occur and can affect measurements considerably. Recall Rule 2: Minimize noise in each projection image. Noise can appear as speckles in slice images, but can be minimized by maximizing the X-ray dose.
There is also non-linear detector noise in projection images that stays in the same position for all projection images. As images form, this noise is reconstructed as circular rings, or so called ring artifacts. Noise in the reference images give the worst ring artifacts because it is amplified each time it is re-used to correct projection images. Noise in the black reference image is more significant than noise in the white reference image, because the signal is less in the black image so the signal-to-noise ratio is also smaller.
Ring artifacts are stronger nearer the axis of rotation because fewer reference image pixels are used. They can be minimized by averaging many frames when collecting black and white reference images.
Beam hardening is the self-filtering of the X-rays by the sample, so the X-rays have a higher energy inside the sample, and are therefore more penetrating. Because of this, the measured X-ray linear attenuation is lower inside the sample than at the edges, giving rise to beam-hardening, or cupping artifacts.
Beam hardening can be reduced by pre-filtering the X-ray beam (placing a filter over the output window of the X-ray source – Rule 3: Use filters to correct beam hardening.). Using beam hardening correction filters in the CT software can also correct it to some extent. This works best with single-material samples.
Rule 5 is use the full range of the detector. Higher dynamic-range detectors (more bits) help detect small differences in intensity due to low-density materials, such as plastics, in the presence of high-density materials, such as metal.
Streak artifacts are caused by beam hardening or lack of penetration of the X-ray beam through the sample. Lack of penetration can be solved by increasing the X-ray energy – unless of course you are at the maximum for the system.
Streak artifacts in this razor blade scan result from the different densities between blades and holder.
Streak artifacts can be reduced by filtering the beam, applying a beam-hardening correction, or by using a detector with a high dynamic range. Scattered radiation can be reduced by collimating the X-ray beam and the detector to detect only those X-rays that travel in a straight line from the source to the detector.
Breaking the Rules
Following the rules will give the best-possible CT results for full metrology purposes, and is general good practice. However, if you need qualitative information (i.e. is a crack present or not, what is the sample’s porosity, how many densities are present, is an electrical connection made or broken), or the information needed will not be affected by the artifacts, then rules can be broken.
Here is a list detailing role by role how each one can be broken:
- Rule 1: Penetrate the sample at all angles
- Rule 6: Keep the object in the field of view.
If the material not in view is relatively uniform in shape and homogeneous, then it will only add a small ring at the edge. Internal features will be easily visible and usable, allowing operators to zoom in and see more detail. If the feature being analyzed is near the center of the scan, or if the size of the feature is much larger than a single pixel, then the number of projections can be reduced to speed up the scan.
- Rule 2: Minimize the noise in each projection image
If time is pressing, reducing the noise in the black and white reference images will reduce the noise significantly without greatly increasing the length of the whole scan.
- Rule 4: Always take projections over 360°
If the object has a very long path that is difficult to penetrate but is narrow so that the lack of penetration only occupies a few projections (<10 to 12), then these projections can be replaced with a corrected image with the sample not present. This adds no attenuation and so it does not affect the reconstruction. However, it is advisable to take some extra projections to compensate.
- Rule 5: Use the whole dynamic range of the detector
When inspecting very low-density specimens, very low energy X-rays give good contrast in the images. To fill the dynamic range requires long exposures. Halving the exposure only loses one bit of information, but it saves a lot of time.
Achivieving Quality Results
The rules of CT come out of the theory of CT reconstruction. Following the rules means you will most likely get the best-quality CT data, but sometimes breaking the rules can save lots of time without compromising image quality too much.
The metrology applications involving X-ray µCT are varied and growing. They include inspecting for intersecting holes in diesel engines; fit-and-finish requirements for drink bottle tips; comparing CT data to CAD models; and dimensioning parts within CT scans, i.e., inventing virtual CMM probes. X-rays are not for special occasions; rather they are tools for everyday use.
Nikon Metrology Inc.
Brighton, MI
https://industry.nikon.com/
Explore the March 2013 Issue
Check out more from this issue and find your next story to read.
Latest from Today's Medical Developments
- US Department of Labor Recognizes UNITED GRINDING North America as Apprenticeship Ambassador
- Navigating today’s supply chain
- Fed’s soft landing may ignite manufacturing technology market growth
- Platinum Tooling named North American distributor for Dunner
- Bridging the Skills Gap: A Solution for Today’s Labor Shortage
- Machine Solutions acquires Alpine Laser LLC
- OSG USA’s PHOENIX PFDC indexable face mill cutter & inserts
- IMTS 2024 Booth Tour: Fagor Automation Corp.